Timing
Table 15. B SWP Timing

Calibration Procedure For Oscilloscope, Os-288/G (Type 2465B) And Types 2465B Opt46 And 2467B Page Navigation
  10    11    12    13    14  15  16    17    18    19    20  
img
. . . .
TB 9-6625-2295-24
(6) Adjust  POSITION  control to align 2d time marker with 2d vertical graticule
line. Rotate oscilloscope calibrator knob, below EDIT FIELD pushbutton, to align 10th
If oscilloscope calibrator err readout
time marker with 10th vertical graticule line.
indication is not within limits specified in first row of table 14, perform b below.
(7) Repeat technique of (2) through (6) above for remaining settings in table 14. If
oscilloscope calibrator err readout indication is not within limits specified in table 14,
perform b below.
Table 14. A SWP Timing
Test instrument
Oscilloscope calibrator
A SWP SEC/DIV
MARKER mode
err display
setting
output
indication (%)
5 ns
5  ns
1.45
10 ns
10  ns
1.45
20 ns
20  ns
1.45
50 ns
50  ns
1.45
100 ns
100  ns
1.45
200 ns
200  ns
1.45
500 ns
500  ns
1.45
1 s
1
1.45
s
2 s
2
1.45
s
5 s
5
1.45
s
10 s
10
1.45
s
20 s
20
1.45
s
50 s
50
1.45
s
100 s
100
1.45
s
200 s
200
1.45
s
500 s
500
1.45
s
1 ms
1  ms
1.45
2 ms
2  ms
1.45
5 ms
5  ms
1.45
10 ms
10  ms
1.45
20 ms
20  ms
1.45
50 ms
50  ms
1.45
100 ms
0.1 s
1.45
200 ms
0.2 s
1.95
500 ms
0.5 s
1.95
500 ms1
50  ms
1.95
1
Set TI X10 MAG to on.
(8) Press Xl0 MAG pushbutton to off.
(9) Set A SWP SEC/DIV switch to 10 ns and B SWP SEC/DIV switch to 5 ns
(knob in) for B SWP only.
(10) Set oscilloscope calibrator for a CHAN 1, MARKER mode output of 5 ns.
(11) Adjust TRIGGER LEVEL control for a stable display.
(12) Adjust  POSITION  control to align 2d time marker with 2d vertical graticule
line. Rotate oscilloscope calibrator knob, below EDIT FIELD pushbutton, to align 10th
15