TB 9-6625-2295-24
(28) Adjust ' control for a 't readout indication near 10.00 ns to superimpose the
two time markers. If 't readout indication is not within limits specified in first row of table
17 with the two time markers superimposed, perform b below.
(29) Repeat technique of (28) above for 't readout indications listed in table 17. If 't
readout indications are not within specified limits in table 17 with the two time markers
superimposed, perform b below.
Test instrument
't readout indications limits
't readout
indications (ns)
Min (ns)
Max (ns)
10.00
10.32
9.68
0.0
0.30
0.30
10
9.68
10.32
20
19.64
20.36
30
29.62
30.38
40
39.58
40.42
50
49.56
50.44
60
59.52
60.48
70
69.50
70.50
80
79.46
80.54
(30) Set A SWP SEC/DIV switch to 20 ns and B SWP SEC/DIV switch to 5 ns
(knob out).
(31) Press X10 MAG pushbutton on and 't pushbutton off for DLY readout.
(32) Set oscilloscope calibrator for a CHAN 1, MARKER mode output of 20 ns.
(33) Adjust 'REF OR DLY POS control for DLY readout of 21.25 ns and
POSITION control to position leading edge of 2d time marker near graticule center.
(34) Set oscilloscope calibrator for a CHAN 1, MARKER mode output of 5 ns.
(35) Press 't pushbutton to obtain a 't display and push in SEC/DIV knob for B
SWP only.
(36) Adjust ' control for a 't readout indication near 20.00 ns to superimpose the
two time markers. If 't readout indication is not within limits specified in first row of table
18 with the two time markers superimposed, perform b below.
(37) Repeat technique of (36) above for 't readout indications listed in table 18. If 't
readout indications are not within specified limits in table 18 with the two time markers
superimposed, perform b below.
Test instrument
't readout indications limits
't readout
indications (ns)
Min (ns)
Max (ns)
20.00
20.45
19.55
20.00
19.55
20.45
160.00
159.15
160.85
18