TB 9-6625-2295-24
time marker with 10th vertical graticule line. If oscilloscope calibrator err readout
indication is not within limits specified in first row of table 15, perform b below.
(13) Repeat technique of (9) through (12) above for remaining settings in table 15. If
oscilloscope calibrator err readout indication is not within limits specified in table 15,
perform b below.
Test instrument
Oscilloscope calibrator
A SWP SEC/DIV B SWP SEC/DIV
MARKER mode
err display
setting
setting
output
indication (%)
10 ns
5 ns
5 ns
1.75
20 ns
10 ns
10 ns
1.75
50 ns
20 ns
20 ns
1.75
100 ns
50 ns
50 ns
1.75
200 ns
100 ns
100 ns
1.75
500 ns
200 ns
200 ns
1.75
1 s
500 ns
500 ns
1.75
2 s
1 s
1 s
1.75
5 s
2 s
2 s
1.75
10 s
5 s
5 s
1.75
20 s
10 s
10 s
1.75
50 s
20 s
20 s
1.75
100 s
50 s
50 s
1.75
200 s
100 s
100 s
1.75
500 s
200 s
200 s
1.75
1 ms
500 s
500 s
1.75
2 ms
1 ms
1 ms
1.75
5 ms
2 ms
2 ms
1.75
10 ms
5 ms
5 ms
1.75
20 ms
10 ms
10 ms
1.75
50 ms
20 ms
20 ms
1.75
100 ms
50 ms
50 ms
1.75
1
5 ms
2 ms
2 ms
1.75
(50 ms displayed)
(20 ms displayed)
Set TI X10 MAG to on.
1
(14) Press X10 MAG pushbutton off.
(15) Press 't pushbutton on (cursors on CRT).
(16) Set oscilloscope calibrator for a CHAN 1, MARKER mode output of 5 ns.
(17) Set A SWP SEC/DIV switch to 5 ns (knobs locked).
(18) Adjust ' REF OR DLY POS control to align one cursor with 2d time marker and
' control to align other cursor with 10th time marker. If 't readout indication is not within
limits specified in first row of table 16, perform b below.
(19) Repeat technique of (16) through (18) above for remaining settings in table 16.
If 't readout indication is not within limits specified in table 16, perform b below.
16