TB 9-6625-2330-24
(3) Configure measuring receiver to measure RF power. Using measuring receiver
and RF power measurement techniques, measuring receiver will indicate within limits
(4) Repeat (1) (a) and (3) above for remaining frequencies listed in table 9.
Measuring receiver
Test instrument
DATA ENTRY
Indications (dB)
Min
Max
0.200
MHz
11.5
14.5
0.500
MHz
11.5
14.5
1
MHz
11.5
14.5
3
MHz
11.5
14.5
10
MHz
11.5
14.5
30
MHz
11.5
14.5
100
MHz
11.5
14.5
200
MHz
11.5
14.5
300
MHz
11.5
14.5
400
MHz
11.5
14.5
500
MHz
11.5
14.5
600
MHz
11.5
14.5
700
MHz
11.5
14.5
800
MHz
11.5
14.5
900
MHz
11.5
14.5
1000
MHz
10.5
15.5
1100
MHz
10.5
15.5
1200
MHz
10.5
15.5
1300
MHz
10.5
15.5
1400
MHz
10.5
15.5
1500
MHz
10.5
15.5
1600
MHz
10.5
15.5
1700
MHz
10.5
15.5
1800
MHz
10.5
15.5
1900
MHz
10.5
15.5
2000
MHz
10.5
15.5
(5) Press TI FUNCTION-RF ON pushbutton to off (red light extinguished).
b. Adjustments. No adjustments can be made.
a. Performance Check
NOTE
If necessary, perform measuring receiver and sensor module
ZERO and CALIBRATE.