Table 14. Spurious Signals.
Figure 2. Pulse Modulation Risetime Hookup.

Calibration Procedure For Signal Generator Sg-1207A/U (, Model Psg2400L) Page Navigation
  11    12    13    14    15  16  17    18    19    20    21  
img
.
TB 9-6625-2330-24
Figure 1. Pulse Modulation On/Off Ratio Hookup.
(3) Press TI pushbuttons as listed in (a) through (e) below:
(a) FUNCTION-CARR FREQ.
(b) DATA ENTRY- 1 GHz.
(c) FUNCTION-CARR LEVEL.
(d) DATA ENTRY- 10 dBm.
(e) FUNCTION-RF ON to on (red light lit).
(4) Configure measuring receiver spectrum analyzer as outlined in (a) through (d)
below:
(a) Preset.
(b) Ref Level to 10 dBm.
(c) Center Frequency to 1 GHz.
(d) Span to 1.5 MHz.
(5) Now configure measuring receiver spectrum analyzer as outlined in (a) through
(j) below:
(a) Resolution bandwidth to 100 kHz.
(b) Video bandwidth to 1 kHz.
(c) Marker to on
(d) Span to 0 Hz.
(e) Sweep Time to 30 ms.
(f) Trigger video to -10 dBm.
(g) Marker to on.
(h) Using rotary knob, adjust marker to top of square wave.
(i) Marker delta to on.
(j)
Using rotary knob, adjust delta marker to bottom of square wave.
16