TB 9-6625-2340-24
(28) Find the largest positive deviation recorded in table 13 and record value in
(29) Find the largest negative deviation recorded in table 13 and record value in
0 to 70 dBm
80 dBm to 120 dBm
(2 dB steps)
(0.1 dB steps)
(10 dB steps)
(10 dB steps)
(A) +
(B) -
(C) +
(D) -
(E) +
(F) -
(G) +
(H) -
(30) Add the values in columns A, E, and G. If total sum exceeds 0.6 dB, perform b below.
(31) Add the values in columns B, F and H. If total sum exceeds 0.6 dB, perform b below.
(32) Add the values in columns A, C, E, and G. If total sum exceeds 1.0 dB, perform
b below.
(33) Add the values in columns B, D, F, and H. If total sum exceeds 1.0 dB, perform
b below.
b. Adjustments
OUTPUT from TI RF INPUT.
(1) Disconnect synthesizer/level generator 50
(2) Set TI LINE switch to STANDBY.
(3) Remove TI top cover.
(4) Remove cables attached to the A3 digital storage cage.
(5) Remove digital storage cage cover.
(6) Reconnect cables to digital storage cage removed in (4) above.
(7) Set TI LINE switch to ON.
(8) Connect TI RF INPUT to TI CAL OUTPUT.
(9) Press TI 2-22 GHz key.
(11) Disconnect 7 (violet) cable from A4A1J1 and terminate cable into a 50
load.
(12) Press TI keys as listed in (a) through (i) below:
(a)
(b)
SWEEP SINGLE.
(c)
TRACE A CLEAR-WRITE.
(d)
MARKER NORMAL.
(e)
MARKER .
(f)
SWEEP CONT.
(g)
SHIFT.
(h)
TRACE A BLANK.
(i)
LVL indication, as indicated by TI
annotation, flickers back and forth between .00 and .10 dB.
39