IF Gain Uncertainty
IF Gain Uncertainty - Continued

Calibration Procedure For Spectrum Analyzer, Model 8566B Page Navigation
  36    37    38    39    40  41  42    43    44    45    46  
img
. . . .
TB 9-6625-2340-24
Figure 24. A4A2 board.
(35) Press TI SCALE LOG key.
(36) Connect TI RF INPUT to synthesizer/level generator 50
OUTPUT.
(37) Set synthesizer/level generator for an output frequency of 7.6 MHz at an output
level of 5 dBm.
(38) Adjust A4A3C55 (fig. 25) for a maximum multimeter indication.
Figure 25. A4A3 board.
NOTE
If A4A3C55 is at an extreme of its adjustment range (fully
meshed or unmeshed) then TI must be repaired.
(39) Short A4A3TP7 to A4A3TP8 (fig. 25) and record multimeter indication.
(40) Remove short from between A4A3TP7 and A4A3TP8.