Calibration Procedure For Measuring Receiver Model 8902A Page Navigation
  40    41    42    43    44  45  46    47    48    49  
img
. . . .
TB 9-6625-2343-40
Figure 19. High select filters - equipment connection.
(24) Press TI keys as listed in (a) through (l) below:
(a)
DATA 41.0.
(b)
DATA SPCL.
(c)
DATA 1200.
(d)
DATA MHz.
(e)
DATA 23.1.
(f)
DATA SPCL.
(g)
DATA 24.0.
(h)
DATA SPCL.
(i)
DATA 24.5.
(j)
DATA SPCL.
(k)
DATA 24.6.
(l)
DATA SPCL.
(25) Increase signal generator No. 2 frequency to 800.465 MHz.
(26) Increase signal generator No. 1 frequency to 800.010 MHz and change output
level to -75 dBm.
(27) TI indication will be within limits specified in first row of table 27.
(28) Repeat technique of (26) above changing output to levels listed in table 27. TI
will indicate within limits specified in table 27.