Figure 19. High select filters - equipment connection.
Power Supply

Calibration Procedure For Measuring Receiver Model 8902A Page Navigation
  41    42    43    44    45  46  47    48    49  
img
.
TB 9-6625-2343-40
Table 27. Level Ratio
Signal generator
Test instrument
No. 1
output
Indication limits
level
(dB)
(dBm)
Min
Max
-75
-76.5
-73.5
-65
-66.5
-63.5
-45
-46.5
-43.5
-25
-26.5
-23.5
-10
-11.5
-8.5
(29) Reduce output of signal generator No. 2 and disconnect from equipment setup.
(30) Set signal generator No. 1 for an output frequency of 455 kHz at an output level
of 3 dB.
(31) Press TI keys as listed in (a) through (l) below:
(a)
BLUE.
(b)
INSTR PRESET.
(c)
DATA 100.
(d)
DATA MHz.
(e)
DATA 24.0.
(f)
DATA SPCL.
(g)
DATA 3.1.
(h)
DATA SPCL.
(i)
DATA 24.1.
(j)
DATA SPCL.
DATA 49., MEASUREMENT S, 2.
(k)
(l)
DATA SPCL.
(32) Adjust signal generator No. 1 output level for a TI indication of 2.500.
(33) Decrease signal generator No. 1 output level by 5 dB.
(34) Divide TI signal displayed level by 1.406 and record the result.
(35) Press TI keys as listed in (a) through (h) below:
(a)
DATA 0.390.
(b)
DATA SPCL.
DATA 0.3, MEASUREMENT S, 4, MEASUREMENT S, 4.
(c)
(d)
DATA SPCL.
DATA 0.3, MEASUREMENT S, 30.
(e)
(f)
DATA SPCL.
DATA 49., MEASUREMENT S, 2.
(g)
(h)
DATA SPCL.
(36) Adjust signal generator No. 1 output level for a TI indication of 1.000.
(37) Decrease signal generator No. 1 output level by 5 dB.
(38) Press TI keys as listed in (a) through (d) below:
(a) DATA 0.3, MEASUREMENT S, 4, MEASUREMENT S, 3.
47