Power Supply
Final Procedure

Calibration Procedure For Measuring Receiver Model 8902A Page Navigation
  43    44    45    46    47  48  49  
img
. . . . . .
TB 9-6625-2343-40
(5) Move positive lead from TP5 to TP6 (fig. 20) and verify multimeter indicates
between -4.9 and -5.2 V dc for serial number 2305A to 2616A and between -5.4 and -4.9 V
dc for serial number 2618A and above.
(6) Move positive lead from TP6 to TP7 (fig. 20) and verify multimeter indicates
between 40.5 and 42.5 V dc.
(7) Disconnect multimeter from A10 board.
POWER SUPPLY ADJUST
Figure 20. A10 board.
(8) Connect multimeter to A13TP10 (fig. 21) (positive) and A13TP15 GND (fig. 21)
(negative).
BROWN
BATTERY BACKUP THRESHOLD ADJUST
Figure 21. A13 board.
(9) Measure and record dc voltage as level No. 1.
49