Table 24. IF Range to Range Linearity
High Select Filters

Calibration Procedure For Measuring Receiver Model 8902A Page Navigation
  38    39    40    41    42  43  44    45    46    47    48  
img
. . . . . . . . . .
TB 9-6625-2343-40
(26) Step signal generator No. 1 output frequency down by 1 kHz. TI will display
within limits specified in first row of table 25.
(27) Repeat technique of (26) above for frequency steps listed in table 25 below. TI
will display within limits specified in first row of table 25.
Table 25. Negative Frequency Drift Error
Signal generator No. 1
Test instrument
output
Indications
frequency
(dB)
(MHz)
Min
Max
29.999
0.05
-0.05
29.998
0.10
-0.10
29.997
0.15
-0.15
29.996
0.20
-0.20
29.995
0.25
-0.25
(28) Set signal generator No. 1 for output frequency to 30 MHz.
(29) Step signal generator No. 1 output frequency up by 1 kHz.
TI will display
within limits specified in first row of table 26.
(30) Repeat technique of (29) above for frequency steps listed in table 26 below. TI
will display within limits specified in first row of table 26.
Table 26. Positive Frequency Drift Error
Signal generator No. 1
Test instrument
output
Indications
frequency
(dB)
(MHz)
Min
Max
30.001
0.05
-0.05
30.002
0.10
-0.10
30.003
0.15
-0.15
30.004
0.20
-0.20
30.005
0.25
-0.25
(31) Reduce all outputs to minimum and disconnect equipment setup.
b. Adjustments. None
17. High Select Filters
a. Performance Check
(1) Press TI keys as listed in (a) and (b) below:
(a) BLUE.
(b) INSTR PRESET.
(2) Connect signal generator No. 1 RF output to TI sensor module.
(3) Set signal generator No. 1 for an output frequency of 455 kHz at an output level of 3 dBm.
(4) Press TI keys as listed in (a) through (h) below:
(a)
DATA 100.
(b)
DATA MHz.
(c)
DATA 24.0.
(d)
DATA SPCL.
44