SWR Test - Continued
Tuned Level RF

Calibration Procedure For Measuring Receiver Model 8902A Page Navigation
  34    35    36    37    38  39  40    41    42    43    44  
img
. . . .
TB 9-6625-2343-40
(28) Press TI keys as listed in (a) and (b) below:
(a) DATA 1.7.
(b) DATA SPCL.
(29) Set spectrum analyzer sweep time to 50 ms, clear/write trace 1, and set trace 1
to maximum hold.
(30) Set signal generator No. 1 for a single sweep of 100 sec with a start frequency of
2.5 MHz and a stop frequency of 1.3 GHz.
(31) When sweep is complete, set spectrum analyzer for a marker peak search.
Marker amplitude displayed on spectrum analyzer will be within limits specified in first
row of table 20.
(32) Repeat technique of (28) through (31) above for remaining TI special function
setting in table 20. Marker amplitude displayed on spectrum analyzer will be within limits
specified in table 20.
Table 20. Ext Sensor Range3 and 14 dB Gain
Test instrument
Spectrum analyzer
Special
Marker amplitude
functions
Displayed limit (dB)
≤-14
1.7
≤-17
1.8
(33) Disconnect TI sensor module from SWR bridge DEVICE UNDER TEST connector.
(34) Connect TI INPUT 50 Ω connector directly to SWR bridge DEVICE UNDER
TEST connector.
(35) Press TI keys as listed in (a) and (b) below:
(a) DATA 1.7.
(b) DATA SPCL.
(36) Set spectrum analyzer sweep time to 50 ms, clear/write trace 1, and set trace 1
to maximum hold.
(37) Set signal generator No. 1 for a single sweep of 100 sec with a start
frequency of 2.5 MHz and a stop frequency of 1.3 GHz.
(38) When sweep is complete, set spectrum analyzer for a marker peak search.
Marker amplitude displayed on spectrum analyzer will be ≤-16.0 dB.
(39) Press spectrum analyzer Preset key.
(40) Set the spectrum analyzer for a start frequency of 3 MHz, stop frequency of 1 GHz
and video bandwidth of 30 kHz.
(41) Disconnect TI INPUT 50 Ω from SWR bridge DEVICE UNDER TEST connector.
(42) Set signal generator No. 1 for a frequency output of 500 MHz at a level of 3 dBm.
(43) Set spectrum analyzer for peak search.
(44) Set spectrum analyzer reference level to marker reference value.
(45) Set spectrum analyzer for peak search.
(46) Activate spectrum analyzer delta marker.
(47) Connect TI INPUT 50 Ω directly to SWR bridge DEVICE UNDER TEST connector.
40


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.