SWR Test
Table 20. Ext Sensor Range3 and 14 dB Gain

Calibration Procedure For Measuring Receiver Model 8902A Page Navigation
  33    34    35    36    37  38  39    40    41    42    43  
img
. . . .
TB 9-6625-2343-40
NOTE
In the tests that follow, ignore error messages displayed by the
measuring receiver.
(11) Set spectrum analyzer sweep time to 50 ms, clear/write trace 1, and set trace 1
to maximum hold.
(12) Set signal generator No. 1 for a single sweep of 100 sec with a start frequency of
2.5 MHz and a stop frequency of 1.3 GHz.
(13) When sweep is complete, set spectrum analyzer for a marker peak search.
Marker amplitude displayed on spectrum analyzer will be ≤-23.0 dB.
(14) Press TI keys as listed in (a) through (f) below:
(a)
MEASUREMENT S (shift).
(b)
TUNED RF LEVEL.
(c)
DATA 1.6.
(d)
DATA SPCL.
(e)
DATA 4.6.
(f)
DATA SPCL.
(15) Set spectrum analyzer for clear/write trace 1, and set trace 1 to maximum hold.
(16) Set signal generator No. 1 for a single sweep of 100 sec with a start frequency of
2.5 MHz and a stop frequency of 1.3 GHz.
(17) When sweep is complete, set spectrum analyzer for a marker peak search.
Marker amplitude displayed on spectrum analyzer will be within limits specified in first
row of table 19.
(18) Repeat technique of (14) (e), (f), and (15) through (17) above for remaining TI
special function setting in table 19. Marker amplitude displayed on spectrum analyzer will
be within limits specified in table 19.
Table 19. Ext Sensor Ranges 1 & 2
Test instrument
Spectrum analyzer
Special function
Marker amplitude
Displayed limit (dB)
≤-17
1.6
≤-17
1.2
(19) Press spectrum analyzer Preset key.
(20) Set the spectrum analyzer for a start frequency of 3 MHz, stop frequency of 1 GHz
and video bandwidth of 30 kHz.
(21) Disconnect TI sensor module from SWR bridge DEVICE UNDER TEST connector.
(22) Set signal generator No. 1 for a frequency output of 500 MHz at a level of -27 dBm.
(23) Set spectrum analyzer for peak search.
(24) Set spectrum analyzer reference level to marker reference value.
(25) Set spectrum analyzer for peak search.
(26) Activate spectrum analyzer delta marker.
(27) Connect TI sensor module to SWR bridge DEVICE UNDER TEST connector.