TB 11-6625-2713-50
(2) Adjust TRIG LEVEL CENTERING ADJ A3A5R6 (fig. 5) for +0.65 volt
(3) Adjust A3A5R82 (fig. 5) for -0.25 volt indication on digital voltmeter (R).
(4) Adjust CRT BIAS ADJ A1A2R12 (fig. 6) for 50 volts indication on digital
(5) Adjust A1A2R24 (fig. 6) for no shift of spot as X10 VERNIER control is pulled
out and pressed in (R).
Figure 6. Horizontal amplifier, test point location.
a. Performance Check
(1) Connect oscillator (A4) to TI EXT HOR input, using cable (B4).
(2) Set both TI SOURCE switches to EXT.
(3) Adjust oscillator frequency for 50 kHz and amplitude for 6 divisions of deflection
on crt.
If
oscillator does not indicate 5 MHz or greater, perform b below.
14