TB 9-5210-219-40
(9) Remove 5.000 inch gage block stack from sine plate.
(10) Wait approximately 2 minutes for TI indication to settle. Press TI HOLD
button. After about 5 seconds, TI will indicate -0.0. Older model TI's will indicate -P0.
NOTE
Do not move TI until display indicates -0.0 or -P0.
(11) Rotate TI 180 degrees on sine plate to same location against sine plate rails. TI
will remain right side up and face away from technician.
Press TI HOLD
(12) Wait approximately 2 minutes for TI indication to settle.
button.
b. TI's with 60 degree range
NOTE
Adjustments to TI will recalculate internal correction factors
which are stored in TI memory. If TI battery is removed,
correction factors calculated in this paragraph will be lost and
TI will revert back to factory determined values.
(2) Use precision level to level sine plate length and width to within 10 arc-sec.
(3) Remove rear cover of TI. Set TI to On.
NOTE
Some older TI's do not have a calibration switch and internal
adjustment cannot be performed.
(4) Refer to figure 4 below. Use a small, blunt object to press TI calibration switch.
CAUTION
Do not move TI sensor shown on left side of TI in figure 4.
(5) Replace TI rear cover. TI will indicate P45.
(6) Place TI on sine plate as shown in figure 1.
(7) Insert a 7.071 inch gage block stack into sine plate.
(8) Wait approximately 2 minutes for TI indication to settle. Press TI HOLD
button. After about 5 seconds, TI will indicate P0.
NOTE
Do not move TI until display indicates P0.
(9) Remove 7.071 inch gage block stack from sine plate.
Press TI HOLD
(10) Wait approximately 2 minutes for TI indication to settle.
button. After about 5 seconds, TI will indicate -P0.
NOTE
Do not move TI until display indicates -P0.