SECTION V. CALIBRATION PROCESS FOR COAXIAL CIRCULATORS
Insertion Loss and Isolation - Continued

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  12    13    14    15    16  17  18    19    20    21    22  
img
. . . . . . . .
TB 9-5985-314-24
NOTE
Clean all connectors with alcohol.
Connect equipment as shown in figure 1, and allow 1 hour for
equipment warm-up.
NOTE
Do not connect TI into equipment setup at this time.
18. Insertion Loss and Isolation
a. Performance Check
(1) Adjust signal generator No. 1 frequency controls to 0.960 GHz and adjust RF
OUTPUT controls for +6 dBm.
NOTE
Use 15% SEARCH MODE on receiver system.
(2) Connect POINT A to POINT B (fig. 1) and establish a reference 0.00 dB on
receiver system at 0.960 GHz.
(3) Connect TI into figure 1 equipment setup as listed in (a) through (c) below:
(a) INPUT PORT to POINT A.
(b) OUTPUT PORT to POINT B.
(c) SIDE PORT to 50 : termination.
(4) Receiver system indication will be 0.5 dB or less.
(5) Adjust signal generator No. 1 and receiver system frequency controls to
frequency settings listed in table 5. Receiver system indication will be 0.5 dB or less.
Table 5. Coaxial Circulator Frequencies
GHz
0.960
1.000
1.030
1.060
1.090
1.100
(6) Repeat (1) and (2) above.
(7) Connect TI into figure 1 equipment setup as listed in (a) through (c) below:
(a) OUTPUT PORT to POINT A.
(b) INPUT PORT to POINT B.
19