SECTION VI. CALIBRATION PROCESS FOR POWER SPLITTERS AND POWER DIVIDERS
Insertion Loss and Output Port Tracking

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  15    16    17    18    19  20  21    22    23    24    25  
img
. . . . .
TB 9-5985-314-24
NOTE
Clean all connectors with alcohol before proceeding with
b below.
b. Connect equipment as shown in figure 4 for TIs with frequency range from 10 MHz
to 18 GHz. Allow 1 hour for equipment warm-up.
NOTE
Equipment shown in figure 5 is for TIs with frequency range
from 18 to 26.5 GHz. Allow 3 hours for equipment warm-up.
NOTE
Equipment shown in figure 6 is for TIs with frequency range
from 26.5 to 40 GHz. Allow 3 hours for equipment warm-up.
NOTE
Do not connect TI into equipment setup at this time.
Figure 4. Power splitter measurement (10 MHz to 18 GHz) - equipment setup.
22. Insertion Loss and Output Port Tracking