TB 9-5985-314-24
a. Performance Check
NOTE
Perform 1 through 4 below for TIs with frequency range listed
1. 10 MHz to 1.9 GHz perform (1) through (11).
2. 2 to 18 GHz perform (12) through (22).
3. 18 to 26.5 GHz perform (23) through (35).
4. 26.5 to 40 GHz perform (36) through (48).
OUTPUT controls for +6 dBm.
NOTE
Use 15% SEARCH MODE on receiver system.
(3) Connect POINT A to POINT B (fig. 4) and establish a 0.00 dB reference on
receiver system at 10 MHz.
(4) Connect TI into 4 equipment setup as listed in (a) through (c) below:
(a) INPUT PORT 1 to POINT A.
(b) OUTPUT PORT 2 to POINT B.
(c) OUTPUT PORT 3 to 50 : termination.
(5) Record receiver system indication in table 6 insertion loss for OUTPUT
PORT 2 column.
remaining test frequency settings listed in table 6. Record receiver system indication in
insertion loss for OUTPUT PORT 2 column in table 6. Insertion loss will be equal to or