Insertion Loss and Isolation - Continued
Figure 4. Power splitter measurement (10 MHz to 18 GHz) - equipment setup.

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  14    15    16    17    18  19  20    21    22    23    24  
img
. . . . . . . . .
TB 9-5985-314-24
(22) Repeat (1) and (2) above.
(23) Connect TI as shown in figure 1, equipment setup as listed in (a) through (c) below:
(a) INPUT PORT to POINT A.
(b) SIDE PORT to POINT B.
(c) OUTPUT PORT to 50 : termination.
(24) Receiver system indication will be 20 dB or greater.
(25) Adjust signal generator No. 1 and receiver system frequency controls to
frequency settings listed in table 5. Receiver system indication will be 20 dB or greater.
b. Adjustments. No adjustments can be made.
19. Final Procedure
a. Deenergize and disconnect all equipment and reinstall protective cover on TI.
b. Annotate and affix DA label/form in accordance with TB 750-25.
SECTION VI
CALIBRATION PROCESS FOR
POWER SPLITTERS AND POWER DIVIDERS
20. Preliminary Instructions
a. The instructions outlined in paragraphs 20 and 21 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning
the calibration.
b. Items of equipment used in this procedure are referenced within the text by
common name.
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
21. Equipment Setup
a. Use Appendix D to determine TI parameters for insertion loss and output tracking.
21