TB 9-5985-314-24
Table 7. Insertion Loss - Output Port Tracking (2 to 18 GHz) Continued
Receiver system indication
OUTPUT
insertion loss
OUTPUT
OUTPUT
PORT
Test
PORT 2
PORT 3
TRACKING
frequency
(dB)
(dB)
(dB)
(GHz)
14.500
15.000
15.500
16.000
16.500
17.000
17.500
18.000
(18) Disconnect TI from equipment setup and repeat (13) and (14) above.
(19) Connect TI into figure 4 equipment setup as listed in (a) through (c) below:
(a) INPUT PORT 1 to POINT A.
(b) OUTPUT PORT 3 to POINT B.
(c) OUTPUT PORT 2 to 50 : termination.
(20) Record receiver system indication in OUTPUT PORT 3 column of table 7.
(21) Adjust signal generator No. 1 and receiver system frequency controls to measure
and record insertion loss for OUTPUT PORT 3 for remaining frequencies listed in table 7.
Insertion loss will be equal to or less than the limits listed in Appendix D.
(22) Algebraically calculate the difference between the values recorded for OUTPUT
PORT 2 and OUTPUT PORT 3 columns of table 7. Record difference in OUTPUT PORT
TRACKING column in table 7. Output tracking will be within the limits specified in
NOTE
Perform (23) through (35) below for TI with frequency range
from 18 to 26.5 GHz.
OUTPUT controls for +3 dBm.
controls for +8 dBm.
26