Table 9. Insertion Loss - Output Port Tracking (26.5 to 40 GHz)
APPENDIX A. DIRECTIONAL COUPLERS TEST INSTRUMENT IDENTIFICATION

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  24    25    26    27    28  29  30    31    32    33    34  
img
. . . . . . . . . .
TB 9-5985-314-24
(45) Connect TI into figure 6 equipment setup as listed in (a) through (c) below:
(a) INPUT PORT 1 to POINT A.
(b) OUTPUT PORT 3 to POINT B.
(c) OUTPUT PORT 2 to 50 : termination.
(46) Record receiver system indication in OUTPUT PORT 3 column of table 9.
(47) Adjust signal generator No. 1 and signal generator No. 2 frequency controls to
remaining settings listed in table 9.  Measure and record insertion loss for OUTPUT
PORT 3 in table 9.
(48) Algebraically calculate the difference between the values recorded in table 9 for
OUTPUT PORT 2 and OUTPUT PORT 3. Record difference in OUTPUT PORT
TRACKING column of table 9. Output port tracking will be within the limits specified in
Appendix D.
b. Adjustments.
Prepare charts (if required) similar to tables 7, 8, and 9 for TI
frequency range.
23. Final Procedure
a. Deenergize and disconnect all equipment and reinstall protective cover on TI.
b. Annotate and affix DA label/form in accordance with TB 750-25.
31/(32 Blank)