TB 9-5985-314-24
(33) Record receiver system indication in OUTPUT PORT 3 column of table 8.
remaining settings listed in table 8. Measure and record insertion loss for OUTPUT
(35) Algebraically calculate the difference between the values recorded in table 8 for
OUTPUT PORT 2 and OUTPUT PORT 3. Record difference in OUTPUT PORT
TRACKING column of table 8. Output tracking will be within the limits specified listed in
NOTE
Perform (36) through (48) below for TI with frequency range
from 26.5 to 40 GHz.
Figure 6. Power splitter measurement (26.5 to 40 GHz) - equipment setup
29