Table 8. Insertion Loss - Output Port Tracking (18 to 26.5 GHz)
Table 9. Insertion Loss - Output Port Tracking (26.5 to 40 GHz)

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  22    23    24    25    26  27  28    29    30    31    32  
img
. . . . . . . . .
TB 9-5985-314-24
(33) Record receiver system indication in OUTPUT PORT 3 column of table 8.
(34) Adjust signal generator No. 1 and signal generator No. 2 frequency controls to
remaining settings listed in table 8.  Measure and record insertion loss for OUTPUT
PORT 3 in table 8. Insertion loss will be approximately equal to value listed in Appendix D.
(35) Algebraically calculate the difference between the values recorded in table 8 for
OUTPUT PORT 2 and OUTPUT PORT 3.  Record difference in OUTPUT PORT
TRACKING column of table 8. Output tracking will be within the limits specified listed in
Appendix D.
NOTE
Perform (36) through (48) below for TI with frequency range
from 26.5 to 40 GHz.
(36) Connect equipment as shown in figure 6.
Figure 6. Power splitter measurement (26.5 to 40 GHz) - equipment setup