• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table 6. Insertion Loss - Output Port Tracking (10 MHz to 1.9 GHz)
Table 7. Insertion Loss - Output Port Tracking (2 to 18 GHz) Continued

Calibration Procedure For Directional Couplers, Vswr Bridges, Coaxial Circulators, Power Splitters, And Power Dividers (10 Mhz To 40 Ghz) (General) Page Navigation
  18    19    20    21    22  23  24    25    26    27    28  
img
. . . . . .
TB 9-5985-314-24
(13) Adjust signal generator No. 1 frequency to 2.000 GHz and adjust RF OUTPUT
controls for +6 dBm.
NOTE
Use 15% SEARCH MODE on receiver system.
(14) Connect POINT A to POINT B (fig. 4) and establish a 0.00 dB reference on
receiver system at 2.000 GHz.
(15) Connect TI into figure 4 equipment setup as listed in (a) through (c) below:
(a) INPUT PORT 1 to POINT A.
(b) OUTPUT PORT 2 to POINT B.
(c) OUTPUT PORT 3 to 50 : termination.
(16) Record receiver system indication in OUTPUT PORT 2 column of table 7.
(17) Adjust signal generator No. 1 and receiver system frequency controls to measure
and record insertion loss for OUTPUT PORT 2 for remaining frequencies listed in table 7.
Insertion loss will be equal to or less than the limits listed in Appendix D.
Table 7. Insertion Loss - Output Port Tracking (2 to 18 GHz)
Receiver system indication
OUTPUT
insertion loss
OUTPUT
PORT
Test
OUTPUT
PORT 2
TRACKING
frequency
PORT 3
(GHz)
(dB)
(dB)
(dB)
2.000
2.500
3.000
3.500
4.000
4.500
5.000
5.500
6.000
6.500
7.000
7.500
8.000
8.500
9.000
9.500
10.000
10.500
11.000
11.500
12.000
12.500
13.000
13.500
14.000
25