TB 9-6625-133-40
(16) Press SAMPLE to initiate the adjustment. Adjustment is complete when the
BUSY legend goes out.
(17) Apply characterized negative voltage to TI for the range that requires
adjustment. Allow time for TI indication to stabilize.
(18) Repeat technique of (13) through (16) above for negative range gain adjustment point.
(19) Repeat technique of (11) through (16) above for remaining TI ranges that require
adjustment.
(20) Exit CAL menu by pressing [QUIT] 2 times.
(21) Disconnect equipment set-up.
c. AC Voltage Adjustments
(1) Set rear panel CALIBRATION switch to ENABLE (if necessary).
NOTE
The cable/adapters used during characterization MUST be
used during performance checks and adjustments.
(2) Connect calibrator OUTPUT HI / LO to TI front INPUT HI / LO with cable
used during ACV characterization.
(3) Press TI keys as listed in (a) through (i) below.
(a)
CLEAR
(b)
[PWR UP DFLT]
(c)
Function: ACV
(d)
CONFIG
(e)
[TSFR] (TSFR mode selected)
(f)
[RESL] , [6]
(g)
CONFIG
(h)
[FILT] [100 HZ]
(i)
ACV
for ranges being adjusted.
(5) Set the TI to the first range requiring adjustment.
(6) Set the calibrator output (displayed) to the zero value for the range being
adjusted (i.e., for 200 V range, set calibrator output to 1 V at 1 kHz).
(8) Press [SET] in the CALIBRATION menu.
(9) Enter the true applied value of the calibrator output using TI numeric keypad,
then press [ENTER].
(10) Press SAMPLE to initiate the adjustment. Range zero adjustment is complete
when the BUSY legend goes out and the calibrated measurement is displayed.
25