TB 9-6625-133-40
TI
Calibrator
Range (V)
Cal Point
Display (V)
True Applied (V)
200 m
Range Zero
10 m
1
200 m
Range Gain LF
100 m
1
1
200 m
Range Gain HF
100 m
60
2
Range Zero
10 m
1
2
Range Gain LF
1
1
Range Gain HF1
2
1
60
20
Range Zero
100 m
1
20
Range Gain LF
10
1
1
20
Range Gain HF
10
60
200
Range Zero
1
1
200
Range Gain LF
100
1
1
200
Range Gain HF
100
60
1000
Range Zero
5
1
1000
Range Gain LF
500
1
Range Gain HF1
1000
500
30
1 HF
gain calibrations are iterative. Repeating the HF calibration may improve the result.
(11) Set the calibrator output (displayed) to the range gain LF value for the range
being adjusted. (i.e., for 200 V range, set calibrator output to 100 V at 1 kHz).
(13) Press [SET] in the CALIBRATION menu.
(14) Enter the true applied value of the calibrator output using TI numeric keypad,
then press [ENTER].
(15) Press SAMPLE to initiate the adjustment. Range gain LF adjustment is
complete when the BUSY legend goes out and the calibrated measurement is displayed.
(16) Set the calibrator output (displayed) to the range gain HF value for the range
being adjusted. (i.e., for 200 V range, set calibrator output to 100 V at 60 kHz).
(18) Press [SET] in the CALIBRATION menu.
(19) Enter the true applied value of the calibrator output using TI numeric keypad,
then press [ENTER].
(20) Press SAMPLE to initiate the adjustment. Range gain HF adjustment is
complete when the BUSY legend goes out and the calibrated measurement is displayed.
Range gain HF is iterative; repeating technique (16) through (20) above may improve
result.
(21) Repeat technique of (5) through (20) above for remaining TI ranges requiring adjustment.
(22) When all ranges and modes requiring adjustment are completed, press CAL to
(23) Set calibrator output to STBY.