Table 12. Resistance Adjustments - Continued.

Calibration Procedure For Reference Multimeter, Models 8508A, 8508A/01 And 8508A/02 Page Navigation
  22    23    24    25    26  27  28    29    30    31    32  
img
.
TB 9-6625-133-40
Table 12. Resistance Adjustments.
TI
Resistance Standards
Range
Mode
Cal Point
Nominal
Test Report
(Ω)
2
Normal
Zero
Short
------
1Ω
2
Normal
Range Gain
2
LoI
Zero
Short
------
1Ω
2
LoI
Range Gain
Tru Ω
2
Zero
Short
------
Tru Ω
1Ω
2
Range Gain
Tru Ω LoI
2
Zero
Short
------
Tru Ω LoI
1Ω
2
Range Gain
20
Normal
Zero
Short
------
10 Ω
20
Normal
Range Gain
20
LoI
Zero
Short
------
10 Ω
20
LoI
Range Gain
Tru Ω
20
Zero
Short
------
Tru Ω
10 Ω
20
Range Gain
Tru Ω LoI
20
Zero
Short
------
Tru Ω LoI
10 Ω
20
Range Gain
200
Normal
Zero
Short
------
100 Ω
200
Normal
Range Gain
200
LoI
Zero
Short
------
100 Ω
200
LoI
Range Gain
Tru Ω
200
Zero
Short
------
Tru Ω
100 Ω
200
Range Gain
Tru Ω LoI
200
Zero
Short
------
Tru Ω LoI
100 Ω
200
Range Gain
2k
Normal
Zero
Short
------
1kΩ
2k
Normal
Range Gain
2k
LoI
Zero
Short
------
1kΩ
2k
LoI
Range Gain
Tru Ω
2k
Zero
Short
------
Tru Ω
1kΩ
2k
Range Gain
Tru Ω LoI
2k
Zero
Short
------
Tru Ω LoI
1kΩ
2k
Range Gain
20 k
Normal
Zero
Short
------
10 k Ω
20 k
Normal
Range Gain
20 k
LoI
Zero
Short
------
10 k Ω
20 k
LoI
Range Gain
Tru Ω
20 k
Zero
Short
------
Tru Ω
10 k Ω
20 k
Range Gain
Tru Ω LoI
20 k
Zero
Short
------
Tru Ω LoI
10 k Ω
20 k
Range Gain
200 k
Normal
Zero
Short
------
100 k Ω
200 k
Normal
Range Gain
200 k
LoI
Zero
Short
------
100 k Ω
200 k
LoI
Range Gain
2M
Normal
Zero
Short
------
1MΩ
2M
Normal
Range Gain
2M
LoI
Zero
Short
------
1MΩ
2M
LoI
Range Gain
28