Table 20. (AB) LF-HF Gain.
Table 22. Meter Volt Gain

Calibration Procedure For Scopemeter Models 190-202 And 190-202/Ar Page Navigation
  17    18    19    20    21  22  23    24  
img
. . . . .
TB 9-6625-134-24
(6) Repeat (3) through (5) for remaining rows in table 21.
Table 21. (AB) Volt Gain.
TI
Calibrator
calibration step
Output Level (V)
CL 0799
5m
CL 0800
12.5 m
CL 0801
25 m
CL 0802
50 m
CL 0803
125 m
CL 0804
250 m
CL 0805
500 m
CL 0806
1.25
CL 0807
2.5
CL 0808
5
CL 0809
12.5
CL 0810
25
CL 0811
50
CL 0812
125
CL 0813
250
f. (AB) Zero
(1) If the display status line does not indicate CL 0852 press F1 PREVIOUS or F2
NEXT for an indication of CL 0852.
(2) Connect as shown in figure 1.
Figure 1. Zero Hook-Up.
(3) Press F3 CALIBRATE and wait for the status line to indicate READY.
(4) Disconnect equipment setup.
g. Meter Volt Gain
(1) If the display status line does not indicate CL 0840 press F1 PREVIOUS or F2
NEXT for an indication of CL 0840.
(2) Connect calibrator OUTPUT V A HI and LO to TI METER V and COM
connectors.
(3) Set calibrator for the first output listed in table 22 and press OPR.