TB 9-6625-134-24
NOTE
Once the CALIBRATE key is press the TI will perform the
Warming Up & Pre-Calibration.
This will take about
30 minutes. Wait for the Ready indication.
(3) The display should show the CAL MODE in upper right corner. The status line
at bottom of the screen should indicate WarmingUp (CL 0200):IDLE or End Precal: Ready.
(4) Ensure all inputs are disconnect (open), and press F3 CALIBRATE and wait
about 30 minutes for the status line to indicate READY.
(5) Press F2 NEXT to select the next calibration step CL 201.
(6) Press F3 CALIBRATE and wait for the status line to indicate READY.
(7) Press F2 NEXT to select the next calibration step CL 0570.
(8) Connect INPUT A to oscilloscope calibrator CHAN 1 using a 50
feed through
termination.
(9) Set oscilloscope calibrator as listed in (a) through (e) below.
(a)
LEVEL SINE.
(b)
500, m, V, ENTER.
(c)
50, MHz, ENTER
(d)
CHANNEL, softkey CHAN1.
(e)
OPR/STBY to OPR.
(10) Press F3 CALIBRATE and wait for the status line to indicate READY.
(11) Set oscilloscope calibrator to STDBY.
b. LF-HF Gain
(1) If the display status line does not indicate CL 0654 press F1 PREVIOUS or F2
NEXT for an indication of CL 0654.
(3) Press F3 CALIBRATE and wait for the status line to indicate READY.
(4) Set oscilloscope calibrator to STBY and press TI F2 NEXT to select the next
(5) Repeat (2) through (4) for remaining rows in table 19.
Oscilloscope Calibrator
TI
calibration step
Function
Output Level
CL 0654
EDGE
500 mV
1k
CL 0480
LEVEL SINE
0.62 V p-p
50 k
CL 0481
LEVEL SINE
0.62 V p-p
221 M
CL 06741
EDGE
500 mV
1k
CL 0482
LEVEL SINE
0.62 V p-p
50 k
CL 0483
LEVEL SINE
0.62 V p-p
221 M
Move connection from INPUT A to INPUT B.
1
22