TB 9-6625-134-24
c. (AB) LF-HF Gain
(1) If the display status line does not indicate CL 0660 press F1 PREVIOUS or
F2 NEXT for an indication of CL 0660.
(2) Connect oscilloscope calibrator CHAN 1 to both TI INPUT A and INPUT B
using a BNC tee (do not use a 50 feed through termination).
(3) Set calibrator for the first output listed in table 20 and press OPR.
(4) Press F3 CALIBRATE and wait for the status line to indicate READY.
(5) Set oscilloscope calibrator to STBY and press TI F2 NEXT to select the next
(6) Repeat (3) through (5) for remaining rows in table 20.
Oscilloscope Calibrator
TI
Output Level
Function
(V)
(kHz)
CL 0660
VOLTAGE, SCOPE Z 1 M
300 m
1
CL 0604
500 m
1
CL 0637
STBY
1
CL 0504
500 m
1
CL 0624
STBY
1
CL 0599
10 m
1
CL 0600
25 m
1
CL 0601
50 m
1
CL 0602
100 m
1
CL 0603
250 m
1
CL 0662
2
1
CL 0605
1
1
CL 0606
2.5
1
CL 0607
5
1
CL 0664
20
1
CL 0608
10
1
CL 0609
25
1
CL 0610
50
1
d. (AB) Position
(1) If the display status line does not indicate CL 0619 press F1 PREVIOUS or
F2 NEXT for an indication of CL 0619.
(2) Remove all connections from INPUT A and INPUT B (open inputs).
(3) Press F3 CALIBRATE and wait for the status line to indicate READY.
e. (AB) Volt Gain
(1) If the display status line does not indicate CL 0799 press F1 PREVIOUS or
F2 NEXT for an indication of CL 0799.
(2) Connect calibrator OUTPUT V A HI and LO to both TI INPUT A and INPUT
B using a BNC tee (do not use a 50 feed through termination).
(3) Set calibrator for the first output listed in table 21 and press OPR.
(4) Press F3 CALIBRATE and wait for the status line to indicate READY.
(5) Set oscilloscope calibrator to STBY and press TI F2 NEXT to select the next
23