Table 3. Sweep Timing
Final Procedure

Calibration Procedure For Time Base Type 7B80 Page Navigation
  2    3    4    5    6  7  8  
img
. . .
TB 9-6625-2056-24
Table 3. Sweep Timing - Continued
Test instrument
Oscilloscope calibrator
TIME/DIV
MARKER
err indication
switch settings
output
(%)
5
ms
5
ms
1.5
20
ms
20
ms
1.5
50
ms
50
ms
1.5
0.1
s
0.1
s
1.5
0.2
s
0.2
s
1.5
0.5
s
0.5
s
1.5
1
s
1
s
4
2
s
2
s
4
5
s
5
s
4
(13) Press and release MAG pushbutton to X10 (out).
(14) Repeat technique of (12) above, except use table 4.
NOTE
Due to bandwidth limitations of vertical amplifier, magnified
10, 20, and 50 ns TIME/DIV switch positions are omitted from
table 4.
Table 4. Magnified Sweep Timing
Test instrument
Oscilloscope calibrator
TIME/DIV
MARKER
err indication
switch settings
output
(%)
s
10
n
2.5
0.1
s
20
ns
2.5
0.2
s
50
ns
2.5
0.5
s
s
2.5
1
0.1
s
s
2.5
2
0.2
s
s
2.5
5
0.5
s
s
2.5
10
1
s
s
2.5
20
2
s
s
2.5
50
5
s
0.1
ms
2.5
10
s
0.2
ms
2.5
20
s
0.5
ms
2.5
50
1
ms
0.1
ms
2.5
2
ms
0.2
ms
2.5
5
ms
0.5
ms
2.5
10
ms
1
ms
2.5
20
ms
2
ms
2.5
50
ms
5
ms
2.5
0.1
s
10
ms
2.5
0.2
s
20
ms
2.5
0.5
s
50
ms
2.5
1
s
0.1
s
5
2
s
0.2
s
5
5
s
0.5
s
5
8