Calibrator Output
Table 4. Sweep Timing Accuracy

Calibration Procedure For Oscilloscope Type 485 Page Navigation
  1    2    3  4  5    6    7    8    9  
img
. . . . . .
TB 9-6625-2118-24
b. Adjustments
(1) Adjust CAL 5V ADJ R2130 (fig. 1) for a 50.00 mV indication on multimeter (R).
(2) Adjust 1 kHz ADJ R2105 (fig. 1) until frequency counter indicates 1.0000 kHz (R).
Figure 1. Test instrument - right side view.
(3