TB 9-6625-2118-24
(10) Set oscilloscope calibrator for a 2 ns output.
(11) Adjust horizontal POSITION control to align second sine wave with second
graticule line. If 10th sine wave is not within 1.5 minor divisions of 10th graticule line and
TI incremental error over center 8 divisions are greater than 0.1 divisions, perform b (4) and
(5) below.
(12) Set TIME/DIV switch to 1 ns.
(13) Set oscilloscope calibrator for a 1 ns output.
NOTE
Reposition CH 1 VOLTS/DIV switch as necessary for best
display.
(14) Adjust horizontal POSITION control to align second sine wave with second
graticule line. If 10th sine wave is not within 1.5 minor divisions of 10th graticule line and
TI incremental error over center 8 divisions is greater than 0.1 division, perform b (6) below.
(15) Set TIME/DIV switch to 10 ns and CH 1 VOLTS/DIV switch to .5.
(16) Set oscilloscope calibrator output to 10 ns markers.
(17) Adjust oscilloscope calibrator for 1 time marker per division.
TI will indicate
within 3% as indicated on oscilloscope calibrator control display.
(18) Repeat technique of (17) above for A and B TIME/DIV switch settings listed in
Test instrument
Oscilloscope calibrator
TIME/DIV
output setting
switch settings
20 ns
20 ns
50 ns
50 ns
.1 s
.1 s
.2 s
.2 s
.5 s
.5 s
1 s
1 s
2 s
2 s
5 s
5 s
10 s
10 s
20 s
20 s
50 s
50 s
.1 ms
.1 ms
.2 ms
.2 ms
.5 ms
.5 ms
1 ms
1 ms
2 ms
2 ms
5 ms
5 ms
10 ms
10 ms
20 ms
20 ms
50 ms
50 ms
.1 s
.1 s
.2 s1
.2 s
.5 s1
.5 s
calibrator control display within 3%.