Table 4. Sweep Timing Accuracy
Adjustments - Continued

Calibration Procedure For Oscilloscope Type 485 Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. . . . . . . . .
TB 9-6625-2118-24
b. Adjustments
(1) Adjust oscilloscope calibrator for 0% on control display and adjust H GAIN R1366
(fig. 1) for 1 marker per division over center 8 divisions (R).
(2) Adjust oscilloscope calibrator for 0% on control display and adjust A CAL ADJ
R1308 (fig. 1) for 1 marker per division over center 8 divisions (R).
(3) Adjust C1190 (fig. 2) for 1 sine wave per division over center 8 divisions (R).