TB 9-6625-2118-24
b. Adjustments
(1) Adjust oscilloscope calibrator for 0% on control display and adjust H GAIN R1366
(fig. 1) for 1 marker per division over center 8 divisions (R).
(2) Adjust oscilloscope calibrator for 0% on control display and adjust A CAL ADJ
R1308 (fig. 1) for 1 marker per division over center 8 divisions (R).
(3) Adjust C1190 (fig. 2) for 1 sine wave per division over center 8 divisions (R).