TB 9-4920-460-24
Table 14. RxDirect Output Continued.
RF signal source
Measuring
TI indications
settings
receiver
(W)
Adjusted power
output
settings
Min
Max
indication
(MHz)
(dBm)
235
+10
21
29
245
+10
21
29
255
+10
21
29
265
+10
21
29
275
+10
21
29
285
+10
21
29
295
+10
21
29
300
+10
21
29
310
+10
21
29
320
+10
21
29
330
+10
21
29
340
+10
21
29
350
+10
21
29
360
+10
21
29
370
+10
21
29
380
+10
21
29
390
+10
21
29
400
+10
21
29
(11) Skip the MB, LOC, VOR, and ILS modes.
(12) Connect RF reference source to measuring receiver using 10 foot cable supplied
with TI.
(13) Set RF reference source for a measured output of -9.5 dBm at 29 MHz.
(14) Disconnect 10 foot cable from measuring receiver and connect it to TI
(15) Press TI ENT then 4 (RxRamp) key on keypad.
(16) Verify TI indicates between 21 and 29 W. If TI does not indicate within limits
specified increase or decrease the VAL= value for a 25 W indication.
(17) Repeat steps technique of (12) through (14) above for remaining values in
NOTE
The left / right keys will change frequencies within selected mode.
(18) Skip the MB, LOC, VOR, and ILS modes.
(19) Set RF reference source for an output of 0 dBm at 120 MHz.
(20) Turn TI off and then back on in COMM mode.
(21) Select 120.000 MHz (118 156 MHz range) on TI and hold down XMIT-MEAS
switch. TI will display 120.00000 MHz 0.5 kHz.
29