TB 9-4920-467-24
(5) Set the calibrator to output 17.000 V dc. If TI reading does not indicate between
16.980 and 17.020 V dc, perform b below.
(6) Set the calibrator to output 19.000 V dc. If TI reading does not indicate between
18.980 and 19.020 V dc, perform b below.
(7) Press the TI ETTS BrainPak left Encoder button to return to the main POD
Alignment screen.
b.
Adjustments
(1) Select and activate Align on the UxValidator.
On-screen instructions are
displayed.
(2) Set the calibrator to output 19.000 V dc.
Allow the circuit to stabilize for
2 minutes.
(3) Select and activate OK on the UxValidator. The TI will measure the voltage.
When the measurement is completed, a second on-screen instruction is displayed.
(4)
Set the calibrator to output 15.000 V dc. Select and activate OK.
(5) When both voltage measurements are complete, the alignment progress bar will
show 100%. The TI will calculate the alignment constants and then enter a live-reading
performance check mode.
(6) Press the TI ETTS BrainPak left Encoder knob to save the alignment
constants.
(7)
13