TB 9-4931-488-24
11. Output Level and Meter Accuracy (All Models Except Those Containing
Option 004)
a. Performance Check
(1) Connect measuring receiver to RF OUT.
NOTE
The maximum output of TIs with option 002 is +13 dBm and
with options 002/003 is +12 dBm.
(2 Set OUTPUT LEVEL switch to +20 dBm and RANGE MHz switch to
512 - 256.
(3) Adjust FREQUENCY TUNE control for a 512 MHz indication on
FREQUENCY MHz display and press COUNTER MODE LOCK ON pushbutton.
(4) Set RF OFF/ON switch to ON.
(5) Select the table (table 5 through table 8) that corresponds to the appropriate
option(s) of the TI. Using the techniques of (3) and (4) above, set OUTPUT LEVEL,
RANGE MHz, and FREQUENCY TUNE switches to settings listed in table selected and
adjust OUTPUT LEVEL vernier control(s) for meter indications as listed in table. If
measuring receiver indications are not as specified, perform b below.
Table 5. Output Meter Accuracy (Standard, Options 001 and H66).
Test instrument
Measuring receiver indications
Meter
OUTPUT LEVEL
Min
Max
switch settings
indications
(dBm)
(dBm)
(dBm)
(dB)
+20
-1
+17.5
+20.5
+20
-7
+11.5
+14.5
+20
-10
+8.5
+11.5
+10
+3
+11.5
+14.5
+10
0
+8.5
+11.5
0
+3
+1.5
+4.5
0
0
-1.5
+1.5
Table 6. Output Meter Accuracy (Option 002).
Test instrument
Measuring receiver indications
25 MHz
512 MHz
1000 MHz
OUTPUT LEVEL
Meter
switch settings
indications
Min
Max
Min
Max
Min
Max
(dBm)
(dB)
(dBm)
(dBm)
(dBm)
(dBm)
(dBm)
(dBm)
+20
-2
+16.5
+19.5
+16.0
+20.0
---
---
+20
-7
+11.5
+14.5
+11.0
+15.0
---
---
+20
-10
+8.5
+11.5
+8.0
+12.0
+7.0
+13.0
+10
+3
+11.5
+14.5
+11.0
+15.0
+10.0
+16.0
+10
0
+8.5
+11.5
+8.0
+12.0
+7.0
+13.0
0
+3
+1.5
+4.5
+1.0
+5.0
0.0
+6.0
0
0
-1.5
+1.5
-2.0
+2.0
-3.0
+3.0