Figure 1. Test Instrument - Top Left Hand Side.
Output Level and Meter Accuracy (All Models Except Those Containing

Calibration Procedure For Signal Generator, Sg1112Viu, Sg1112V2U, And Mis-28707Ty2 Page Navigation
  10    11    12    13    14  15  16    17    18    19    20  
img
. . . .
TB 9-4931-488-24
Figure 2. Test Instrument - Top Back Side.
(7) Adjust AM control for TI meter indications listed in table 13. Multimeter should
indicate as specified. If not, readjust A2R29 (METER DRIVE, fig. 1) for best in-tolerance
condition.
Table 13. Meter Accuracy.
Test instrument
Multimeter indications (V rms)
meter indications
Min
Max
(0 - 10 scale)
7
0.487
0.503
5
0.345
0.362
3
0.205
0.220
10
0.697
0.718
(8) Disconnect multimeter from TI.
(9) Connect measuring receiver power sensor module to RF OUT.
(10) Position controls as listed in (a) through (i) below.
(a) LEVEL VOLTS pushbutton pressed.
(b) AM control fully ccw.
(c) FM switch to OFF.
(d) COUNTER MODE LOCK ON released.
(e) RANGE MHz switch 64-32.
(f) FREQUENCY TUNE control for a FREQUENCY MHz display indication of 50 MHz.
15