TB 9-4931-488-24
Table 10. Output Level Accuracy (Option 002).
Test instrument
Measuring receiver indications
OUTPUT LEVEL
25 MHz
512 MHz
1000 MHz
Min
Max
Min
Max
Min
Max
switch settings (dBm)
(dBm)
(dBm)
(dBm)
(dBm)
(dBm)
(dBm)
-10
-12.0
-8.0
-12.5
-7.5
-13.5
-6.5
-20
-22.0
-18.0
-22.5
-17.5
-23.5
-16.5
-30
-32.0
-28.0
-32.5
-27.5
-33.5
-26.5
-40
-42.0
-38.0
-42.5
-37.5
-43.5
-36.5
-50
-52.5
-47.5
-53.0
-47.0
-54.0
-46.0
-60
-62.5
-57.5
-63.0
-57.0
-64.0
-56.0
-70
-72.5
-67.5
-73.0
-67.0
-74.0
-66.0
-80
-82.5
-77.5
-83.0
-77.0
-84.0
-76.0
-90
-92.5
-87.5
-93.0
-87.0
-94.0
-86.0
-100
-102.5
-97.5
-103.0
-97.0
-104.0
-96.0
-110
-112.5
107.5
-113.0
-107.0
-114.0
-106.0
Table 11. Output Level Accuracy (Option 002 with 003).
Measuring receiver indications
Test instrument
25 and 512 MHz
1000 MHz
OUTPUT LEVEL
switch settings
Min
Max
Min
Max
(dBm)
(dBm)
(dBm)
(dBm)
(dBm)
-10
-13.5
-7.5
-14.0
-6.0
-20
-23.5
-16.5
-24.0
-16.0
-30
-33.5
-27.5
-34.0
-26.0
-40
-43.5
-37.5
-44.0
-36.0
-50
-54.0
-47.0
-54.5
-45.5
-60
-64.0
-57.0
-64.5
-55.5
-70
-74.0
-67.0
-74.5
-65.5
-80
-84.0
-77.0
-84.5
-75.5
-90
-94.0
-87.0
-94.5
-85.5
-100
-104.0
-97.0
-104.5
-95.5
-110
-114.0
-107.0
-114.5
-105.5
Table 12. Output Level Accuracy (Option 003).
Measuring receiver indications
Test instrument
OUTPUT LEVEL
Min
Max
switch settings (dBm)
(dBm)
(dBm)
-10
-12.75
-7.75
-20
-22.75
-17.75
-30
-32.75
-27.75
-40
-42.75
-37.75
-50
-53.25
-47.25
-60
-63.25
-57.25
-70
-73.25
-67.25
-80
-83.25
-77.25
-90
-93.25
-87.25
-100
-103.25
-97.25
-110
-113.25
-107.25
13