Table 5. Output Meter Accuracy (Standard, Options 001 and H66).
Table 10. Output Level Accuracy (Option 002).

Calibration Procedure For Signal Generator, Sg1112Viu, Sg1112V2U, And Mis-28707Ty2 Page Navigation
  7    8    9    10    11  12  13    14    15    16    17  
img
. . . .
TB 9-4931-488-24
Table 7. Output Meter Accuracy (Option 002 with 003).
Test instrument
Measuring receiver indication
Meter
25 and 512 MHz
1000 MHz
OUTPUT LEVEL
switch settings
indications
Min
Max
Min
Max
(dBm)
(dB)
(dBm)
(dBm)
(dBm)
(dBm)
+20
-2
+15.0
+20.0
---
---
+20
-7
+10.0
+15.0
---
---
+20
-10
+6.5
+12.5
---
---
+10
+2
+9.0
+14.0
+8.5
+15.5
+10
0
+7.0
+12.0
+6.5
+13.5
0
+3
0.0
+5.0
-0.5
+6.5
0
0
-3.0
+2.0
-3.5
+3.5
Table 8. Output Meter Accuracy (Option 003).
Test Instrument
Measuring receiver indications
Meter
Min
Max
OUTPUT LEVEL
switch settings (dBm)
Indications (dB)
(dBm)
(dBm)
+20
-2
+15.75
+19.75
+20
-7
+10.75
+14.75
+20
-10
+7.75
+11.75
+10
+3
+10.75
+14.75
+10
0
+7.75
+11.75
0
+3
+0.75
+4.75
0
0
-2.25
+1.75
(6)
Adjust OUTPUT LEVEL vernier control(s) for a 0.0 dBm measuring
receiver indication.
(7) Select the table (table 9 through table 12) that corresponds to the appropriate
option(s) of the TI.  Using the techniques of (2) and (3) above set OUTPUT LEVEL,
RANGE MHz, and FREQUENCY TUNE switches to settings listed in the table selected.
(8) Set measuring receiver to measure Tuned RF Level. The measuring receiver
should indicate as specified.
(9) Release COUNTER MODE LOCK ON pushbutton.
Table 9. Output Level Accuracy (Basic, Option 001 and H66).
Measuring receiver indications
Test instrument
OUTPUT LEVEL
Min
Max
switch settings (dBm)
(dBm)
(dBm)
-10
-12
-8
-20
-22
-18
-30
-32
-28
-40
-42
-38
-50
-52.5
-47.5
-60
-62.5
-57.5
-70
-72.5
-67.5
-80
-82.5
-77.5
-90
-92.5
-87.5
-100
-102.5
-97.5
-110
-112.5
-107.5
12