• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table 7. AM Depth Verification.
Table 10. Low Frequency Level Accuracy Verification.

Calibration Procedure For Rf Reference Source Fluke Models 9640A, 9640A-Std And 9640A-Lpn With Leveling Head Model 9640A-50 Page Navigation
  6    7    8    9    10  11  12    13    14    15    16  
img
. . . . . . .
TB 9-5895-1913-40
(13) Repeat technique of 10 (a) (d) (e) (g), (11) and (12) above for the remaining
modulation rates listed in table 8. TI will indicate within the limits of table 8.
Table 8. FM Rate Verification.
TI
Measuring receiver
Frequency
Modulation
Deviation
Min
Max
(MHz)
rate (kHz)
(kHz)
(kHz)
(kHz)
125
1
300
0.99990
1.00010
1000
300
1000
299.990
300.010
(14) Set the TI controls as listed in (a) through (g) below:
(a) FREQUENCY to 125 MHz.
(b) LEVEL to +13 dBm.
(c) MOD to FM.
(d) MOD RATE to 1 kHz (Sine).
(e) FM DEVIATION to 100 kHz.
(f) OUTPUT to OPER.
(g) MODULATION to OFF.
(15) Allow the measuring receiver to tune to the input signal. When the tuning is
complete, set the TI to MODULATION ON.
(16) Set the measuring receiver for an FM Deviation measurement. Verify that the
measuring receiver FM Deviation indications are within the limits of table 9.
(17) Repeat technique of 14 (a) (d) (e) (g) (15) and (16) above for the remaining
modulation rate listed in table 9. TI will indicate within the limits of table 9.
Table 9. FM Deviation Verification.
TI
Measuring receiver
Frequency
Modulation
Deviation
Min
Max
(MHz)
rate (kHz)
(kHz)
(kHz)
(kHz)
125
1
100
97.0
103.0
125
100
100
97.0
103.0
125
1
300
291.0
309.0
125
200
300
291.0
309.0
1000
1
1000
970
1030
1000
200
1000
970
1030
(18) Set TI output to STBY and disconnect equipment set up.
b. Adjustments. No adjustments can be made.