• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table 7. 2 GHz 18 GHz Power Level Accuracy
Output Level Flatness

Calibration Procedure For Synthesized Signal Generator Model Mg3695B, With Options 1B, 2B, 5, 6, And 28B Page Navigation
  6    7    8    9    10  11  12    13    14    15    16  
img
. . .
TB 9-6625-106-24
Table 8. 20 GHz 40 GHz Power Level Accuracy
TI
Measuring Receiver
Output
Min
Max
(dBm)
(dBm)
(dBm)
0
-1
1
-5
-6
-4
-10
-11
-9
-15
-16
-14
-20
-21
-19
-25
-26
-24
-30
-31
-29
-35
-36
-34
-40
-41
-39
-45
-46
-44
-50
-51
-49
-55
-56
-54
-60
-61
-59
-65
-66
-64
-70
-71
-69
-75
-76
-74
-80
-81
-79
-85
-86
-84
-90
-91
-89
-95
-96
-94
(31) Press TI keys as listed in (a) through (c) below:
(a) System, Reset.
(b) EDIT F1, 50, GHz.
(c) Edit L1, 0, dBm.
(32) Using measuring receiver, in measuring receiver mode, and tuned RF power
measuring techniques, verify that the measuring receiver indicates within the limits listed
in table 9.
Table 9. 50 GHz Power Level Accuracy
TI
Measuring Receiver
Output
Min
Max
(dBm)
(dBm)
(dBm)
0
-1.5
1.5
-5
-6.5
-3.5
-10
-11.5
-8.5
-15
-16.5
-13.5
-20
-21.5
-18.5
-25
-26.5
-23.5
-30
-31.5
-28.5
-35
-36.5
-33.5
-40
-41.5
-38.5
-45
-46.5
-43.5
-50
-51.5
-48.5
-55
-56.5
-53.5
-60
-61.5
-58.5
12