Table 5. 10 MHz 15 MHz Power Level Accuracy
Table 7. 2 GHz 18 GHz Power Level Accuracy

Calibration Procedure For Synthesized Signal Generator Model Mg3695B, With Options 1B, 2B, 5, 6, And 28B Page Navigation
  4    5    6    7    8  9  10    11    12    13    14  
img
. . . . . . . . . .
TB 9-6625-106-24
(18) Set L1 to the next value listed in table 6. Measuring receiver will indicate
within limits listed in table 6.
(19) Repeat technique of (18) above for remaining TI settings listed in table 6. TI will
indicate within limits listed in table 6.
(20) Repeat technique of (16) (b) and (c) through (19) above for frequency of 500 MHz,
600 MHz, and 1 GHz.
Table 6. 60 MHz 1GHz Power Level Accuracy
TI
Measuring Receiver
Output
Min
Max
(dBm)
(dBm)
(dBm)
10
9
11
5
4
6
0
-1
1
-5
-6
-4
-10
-11
-9
-15
-16
-14
-20
-21
-19
-25
-26
-24
-30
-31
-29
-35
-36
-34
-40
-41
-39
-45
-46
-44
-50
-51
-49
-55
-56
-54
-60
-61
-59
-65
-66
-64
-70
-71
-69
-75
-76
-74
-80
-81
-79
-85
-86
-84
-90
-91
-89
-95
-96
-94
(21) Press TI keys as listed in (a) through (c) below:
(a) System, Reset.
(b) EDIT F1, 2, GHz.
(c) Edit L1, 5, dBm.
(22) Using measuring receiver, in measuring receiver mode, and tuned RF power
measuring techniques, verify that the measuring receiver indicates within the limits listed
in table 7.
(23) Set L1 to the next value listed in table 7. Measuring receiver will indicate
within limits listed in table 7.
(24) Repeat technique of (23) above for remaining TI settings listed in table 7.
Measuring receiver will indicate within limits listed in table 7.
(25) Repeat technique of (21) (b) and (c) through (24) above for frequencies of 4, 6, 8,
10, 12, 14, 16, 18 GHz.