Figure 2. Zero Test Setup.
Figure 4. Input Impedance Test.

Calibration Procedure For Differential Amplifier, Model 428 Page Navigation
  1    2    3    4  5  6    7    8    9    10  
img
. . . . . . .
TB 9-6625-170-24
10. Gain and Linearity
a. Performance Check
(1) Connect equipment as shown in figure 3.
Figure 3. Gain Check Setup.
(2) Set the GAIN switch to 10.
(3) Set the GAIN IN/OUT switch to the OUT position.
(4) With the TI COARSE and FINE ZERO controls, adjust for a zero reading on
the multimeter.
(5) Set the calibrator to 5 V dc. Multimeter should read 49.9 to 50.1 V dc. If not,
perform b below.
(6) Repeat step (3) above using the remaining gain switch and calibrator settings in
table 4. TI will read within limits in table 4. If not, perform b below.
Table 4. Gain and Linearity.
Calibrator
TI Output
Min
Max
Gain Step
Output
(V dc)
(V dc)
(V dc)
(V dc)
10
5
50
49.9
50.1
10
4
40
39.90
40.10
10
3
30
29.93
30.07
10
2
20
19.95
20.05
10
1
10
9.98
10.02
20
2.5
50
49.9
50.1
50
1
50
49.9
50.1
100
0.5
50
49.9
50.1
200
0.25
50
49.9
50.1
500
0.1
50
49.9
50.1
1000
0.05
50
49.9
50.1
b. Adjustments. None.