TB 9-6625-170-24
S1
S2
R1
100?
1
R2
4
TI
2
100?
Calibrator
9
1
S3
11
2
= 20 log
∗
(8) Using the larger voltage of steps (6 )and (7), compute the common mode rejection
.
(CMR) (in db) as follows:
(9) CMR will be ≥ 140 db.
b. Adjustments. None.
a. Performance Check
Figure 7. CMR AC Balance Test Setup.
(2) Adjust the function generator for a 20 V pp sine wave at 60 Hz.
(3) Set S1 to 2.
(4) Note the peak-to-peak signal display on the oscilloscope as e.
= 20 log
∗
(5) Compute
the
common
mode
rejection
(CMR)
(in
db)
as
follows:
.