• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Final Procedure
Table 5. Adjustment with Power Range Switch Set to 15.

Calibration Procedure For Rf Directional Thruline Wattmeter, Model 4304A With 4304A-1 Element Page Navigation
  5    6    7    8    9  10  11    12  
img
. .
TB 9-6625-187-24
APPENDIX A
PERFORMANCE CHECK TABLES AND ADJUSTMENTS WORKSHEET
Table 3. Measured Values Above 100 MHz.
TI
Power
Tolerance
Difference
TI
Frequency
Low Pass
Power
Meter
Power Meter and
%
reading
(MHz)
Filter
range
reading
TI reading
FS
watts
switch
(watts)
110
TLC125-6EF1
5
5.0
6
0.3
15
15.0
6
0.9
50
50.0
6
3.0
150
100.0
6
9.0
100.01
500
6
30.0
150
TLC200-6EF1
5
5.0
6
0.3
15
15.0
6
0.9
50
50.0
6
3.0
150
100.0
6
9.0
100.0 1
500
6
30.0
225
TLC316-6EF1
5
5.0
6
0.3
15
15.0
6
0.9
50
50.0
6
3.0
150
100.0
6
9.0
100.0 1
500
6
30.0
400
TLC450-6EF1
5
5.0
6
0.3
15
15.0
6
0.9
50
50.0
6
3.0
150
100.0
6
9.0
100.0 1
500
6
30.0
800
TLS1225-5EF1
5
5.0
7
0.35
15
15.0
7
1.05
50
50.0
7
3.5
150
100.0
7
10.5
500
100.0
7
35.0
1000
TLS1225-5EF1
5
5.0
7
0.35
15
15.0
7
1.05
50
50.0
7
3.5
150
100.0
7
10.5
After reading is taken, go to standby and change Low pass filter (if required) before proceeding.