Figure 12. Video Processor - Adjustment and Test Point Locations.
Frequency Accuracy - Continued

Calibration Procedure For Spectrum Analyzer An/Usm-489(V)1 (Type 492 W/Options 01, 02, 03, 08, And 21) Page Navigation
  21    22    23    24    25  26  27    28    29    30    31  
img
. .
TB 9-6625-2134-24
(164) Press 2 dB/DIV VERTICAL DISPLAY pushbutton to on and adjust
REFERENCE LEVEL control until signal amplitude is half screen. If TI has option 01,
adjust PEAKING control for maximum response.
(165) Adjust synthesized signal generator for a 5 to 20 GHz sweep and sweep time
for 99 s.
(166) Press DIGITAL STORAGE VIEW A and VIEW B pushbuttons to on.
(167) Adjust TIME/DIV switch until display is solid (no breaks in digitized display
(fig. 13).
Figure 13. Baseline Leveling - Typical Display Response
Before and After Adjustments.
(168) Press DIGITAL STORAGE MAX HOLD and SAVE A pushbuttons to on.
Note position of display and record.